2902049 Phoenix Contact Electrical Measuring & Monitoring Instruments
Part#:2902049
Manufacturer:Phoenix Contact
Category:Electrical Measuring & Monitoring Instruments
Product Lifecycle:Active
Get a quick Quote!
Configurable temperature transducer with plug-in connection technology for connecting 2, 3, and 4-conductor resistance thermometers and resistance-type sensors. Configurable via DIP switch or software. Screw connection technology, standard configuration
About the manufacturer: Phoenix Contact is one of the manufacturers that offers Electrical Measuring & Monitoring Instruments that you can order at Power Resources International site. Phoenix Contact manufactures industrial electrical and electronic technology products that offer power, protection, connectivity, and automation for systems and equipment.
Check out the 2902049 Phoenix Contact price in the box on the right side or send a quick quote.
Product Details
- Width 6.2 mm
- Permissible humidity (operation) 5 % ... 95 % (non-condensing)
- Configurable/programmable Yes
- Voltage output signal 0 V ... 5 V (via DIP switch)
- Load/output load current output ≤ 600 Ω (at 20 mA)
- Torque 0.5 Nm ... 0.6 Nm
- Conductor cross section AWG 24 ... 12 (flexible)
- Degree of pollution 2
- Protective circuit Transient protection
- Mounting position any
- Noise emission EN 61000-6-4
- Fire protection for rail vehicles (DIN EN 45545-2) R22 HL 1 - HL 2
- REACh SVHC Lead 7439-92-1
- ECLASS 4.0 27200200
- ECLASS 9.0 27210129
- ECLASS 7.0 27200206
- UNSPSC 7.0901 39121008
- Height 109.81 mm
- Ambient temperature (storage/transport) -40 °C ... 85 °C
- Noise immunity EN 61000-6-2 When being exposed to interference, there may be minimal deviations.
- Linear resistance measuring range 0 Ω ... 4000 Ω (Minimum measuring span: 10% of the selected measuring range)
- Connection technology 2-, 3-, 4-conductor
- Current output signal 0 mA ... 20 mA (via DIP switch)
- Test voltage, input/output/supply 3 kV (50 Hz, 1 min.)
- Overvoltage category II
- Assembly instructions The T connector can be used to bridge the supply voltage. It can be snapped onto a 35 mm DIN rail according to EN 60715.
- Standards/regulations EN 61000-4-6
- Certificate DNV GL 14445-15HH
- ECLASS 11.0 27210129
- ECLASS 6.0 27200200
- Temperature measuring range -200 °C ... 850 °C (Range depends on sensor type, range can be set freely via software or in increments from -150°C to 850°C via DIP switches)
- Load/output load voltage output ≥ 10 kΩ
- Transmission error resistance thermometer 0.1 % * 350 K / set measuring range; 0.1 % > 350 K (Pt/Ni)
- Short-circuit current < 31.5 mA
- Supply voltage range 9.6 V DC ... 30 V DC (The DIN rail bus connector (ME 6,2 TBUS-2 1,5/5-ST-3,81 GN, Order No. 2869728) can be used to bridge the supply voltage. It can be snapped onto a 35 mm DIN rail according to EN 60715))
- Screw thread M3
- Identification C, EMC2
- Max. output current 24.6 mA
- Max. output voltage approx. 12.3 V
- Connection method Screw connection
- UL, USA/Canada UL 508 Listed
- ECLASS 5.0 27200200
- ETIM 2.0 EC001446
- ETIM 7.0 EC002919
- UNSPSC 6.01 30211506
- UNSPSC 21.0 41112105
- Utilization restriction EMC: class A product, see manufacturer's declaration in the download area
- Ambient temperature (operation) -40 °C ... 70 °C
- Sensor types (RTD) that can be used Pt, Ni, Cu sensors
- Ripple < 10 mVrms
- Power consumption ≤ 850 mW (at IOUT = 20 mA, 9.6 V DC, 600 Ω load)
- Maximum temperature coefficient 0.01 %/K
- Rated insulation voltage 300 V (effective)
- Designation Shipbuilding approval
- China RoHS Environmentally Friendly Use Period = 50 years
- ECLASS 4.1 27200200
- ETIM 6.0 EC002919
- UNSPSC 11 39121008
- UNSPSC 19.0 41112105
- Typical current consumption 32 mA (24 V DC)
- Conductor cross section solid 0.2 mm² ... 1.5 mm² (with ferrule)
- Electromagnetic compatibility Conformance with EMC directive
- Fire protection for rail vehicles (DIN EN 45545-2) R23 HL 1 - HL 2
- Typical deviation from the measuring range final value 0.07 %
- UNSPSC 12.01 39121008
- UNSPSC 20.0 41112105
- Depth 119.2 mm
- Degree of protection IP20 (not assessed by UL)
- Number of outputs 1
- Electrical isolation Reinforced insulation in accordance with IEC 61010-1
- ATEX II 3 G Ex nA IIC T4 Gc X
- ECLASS 10.0.1 27210129
- ECLASS 5.1 27200200
- ETIM 3.0 EC001446
- UNSPSC 13.2 41112105
- Sensor input current approx. 200 µA
- Stripping length 10 mm
- Conductor cross section flexible 0.14 mm² ... 2.5 mm²
- Transmission error resistance-type sensor 2 Ω
- Housing material PBT
- Fire protection for rail vehicles (DIN EN 45545-2) R24 HL 1 - HL 2
- Conformance CE-compliant
- ETIM 4.0 EC001446
- UNSPSC 18.0 41112105
Product condition
New Factory Sealed
New Factory Sealed - New condition, never used, original OEM part with original packaging and factory seal.
Manufacturer warranty